Case Report
 
Simplified diagnostic-surgical stent for accuracy of implant placement and parallelism
Abdul Habeeb Bin Mohsin1, Husna Khan2, Mohd. Aijaz Ahmed3, M. Mahesh Kumar4, Sarah Samee5
1MDS, Assistant Professor, Department of Prosthodontics, Sri Sai College of Dental Surgery, Vikarabad, India.
2BDS, Dental Surgeon, Mahavir Dental Hospital, Hyderabad, India.
3(MDS), Post Graduate Student, Department of Periodontics, Government Dental College, Vijayawada, India.
4(MDS), Post Graduate Student, Department of Periodontics, Sri Sai College of Dental Surgery, Vikarabad, India.
5BDS, Dental Surgeon, BMS Dental Hospital, Hyderabad, India.

Article ID: 100023CRINTAM2016
doi:10.5348/crint-2016-23-CR-4

Address correspondence to:
Dr. Abdul Habeeb Bin Mohsin
23-2-360/3 Moghal Pura
Hyderabad 500002
Telangana
India

Access full text article on other devices

  Access PDF of article on other devices

[HTML Abstract]   [PDF Full Text] [Print This Article]
[Similar article in Pumed] [Similar article in Google Scholar]

How to cite this article:
Mohsin AHB, Khan H, Ahmed MA, Kumar MM, Samee S. Simplified diagnostic-surgical stent for accuracy of implant placement and parallelism. Case Rep Int 2016;5:13–17.


Abstract
Introduction: This article describes a method of evaluation of available alveolar bone dimensions and guided implant placement in a patient with the fabrication of diagnostic-surgical implant template.
Case Report: We present a case of 37-year-old female patient with a missing tooth to be replaced with dental implant using guiding template. The diagnostic template was transformed into a surgical template for implant surgery. The method described is simple, enabling the practitioner to perform radiographic diagnosis and placing the implant precisely using surgical template. This helps in treatment planning and to assess available alveolar bone at the site of planned implant insertion.
Conclusion: This method is easy to follow; significant time and cost savings can therefore be realized by using this template to perform dental implant osteotomy. It provides adequate accuracy in terms of implant position, angulation and predictable treatment outcome.

Keywords: Implant, Imaging, Position, Stent


Introduction

Radiographs used for the evaluation of the available bone dimensions for implant insertion is generally recognized procedure. For the posterior maxillary/mandibular regions a diagnostic radiograph of the jaws is mandatory if implant placement is considered at these sites in order to avoid possible damage to the inferior alveolar nerve or maxillary sinus penetration.

Taking panoramic radiographs is common practice because of the radiographic information it conveys, and its relatively low radiation exposure. However, in some cases periapical radiographs can also be used for assessing the bone dimensions.

Procedure of having a metal ball of known diameter incorporated into an acrylic template (stent) fabricated on diagnostic model of the patient. The use of diagnostic and surgical templates for implant placement may be helpful. For the advantage of transferring radiographic information diagnostic templates are modified to the surgical template [1].

Predictable implant-supported restorations require a determination of the final prosthesis in the treatment planning stage. A radiographic-surgical template can illustrate the final result and the procedures needed to accomplish it.


Case Report

A 37-year-old female patient reported to the Department of Prosthodontics and Implantology with the chief complaint of missing tooth in lower left back teeth region since nine months and wants her tooth to be replaced. History of present illness revealed that there was a tooth decay and pain due to which she got her tooth extraction done nine months back. There is no relevant medical history. Intraoral examination revealed that there is missing lower left first molar with adequate mesiodistal, buccolingual and cervico-occlusal space present (Figure 1). Radiographically there is no pathology detected. Patient was advised to go for surgical profile; the results suggest that all the readings are in the normal range.

The case as discussed was planned so as to place one endosteal implant in the left mandibular first molar area with the help of an implant template. This template was first used to determine the position in a panoramic X-ray and then the same was transformed, used as a surgical template for the placement of implant. Single tooth implant placement planned with a technique to allow precise implant placement with customized diagnostic/surgical template, internal precision guide and guide wire to facilitate improved pre-surgical implant diagnostics and prosthetically directed implant placement.

Accurate impressions were made using an irreversible hydrocolloid impression material (Tropicalgin by Zhermack). Impressions were poured with Type III dental stone (Figure 2). After the stone casts have adequately hardened, the most optimal position for the implant determined. Cast marked with a pencil. The processed acrylic tooth with clear auto polymerizing acrylic resin (DPI-RR Cold Cure) template is modified by drilling a hole and placing the metal ball through the occlusal surface (Figure 3) and placing the stainless steel wire on buccal side of the template (Figure 4). The position and orientation of the proposed implant is identified by the radiopaque metal ball within the proposed restoration. Incorporation of wires placed on the buccal surface of the template allows identification of the preferred implant site and parallelism (Figure 5). After removing the metal ball, the buccal wire remains a constant reference point to be used throughout the implant surgery.

The diagnostic template was transformed into a surgical template for implant surgery (Figure 6) and (Figure 7). A guide hole of 5 mm in diameter was drilled where the metal ball was previously placed as it is the optimum implant site. The surgical template helped to identify the position for implant placement. The safe-distance measured from the mandibular canal. The mesiodistal and buccolingual width was measured with the help of a bone gauge. This led to the choice of implant of 4.4 mm in diameter and 10 mm in length.

The patient was premedicated with antibiotics (novamox cipla 1 g, one hour prior to surgery). Surgery was performed under local anesthesia. The template was placed in the mouth and a crestal mark was placed with a pilot drill. Supracrestal incisions were made, buccal and lingual full thickness mucoperiosteal flaps were raised. Using the surgical template, LASAK -Implant BioAccel- Tapered implant D4.4/L10 mm was placed at tooth #36 location (Figure 8). Surgical cover screws were placed and the flaps were approximated with primary closure (Figure 9). Postoperative panoramic radiograph was taken (Figure 10).


Cursor on image to zoom/Click text to open image
Figure 1: Preoperative view showing missing lower left first molar.



Cursor on image to zoom/Click text to open image
Figure 2: Diagnostic study model.



Cursor on image to zoom/Click text to open image
Figure 3: Diagnostic template with metal ball.



Cursor on image to zoom/Click text to open image
Figure 4: Buccal wire for parallelism.



Cursor on image to zoom/Click text to open image
Figure 5: Radiopaque metal ball and buccal wire in panoramic X-ray.



Cursor on image to zoom/Click text to open image
Figure 6: Template transformed into surgical template.



Cursor on image to zoom/Click text to open image
Figure 7: Surgical template intra-orally.



Cursor on image to zoom/Click text to open image
Figure 8: Final Implant placement.



Cursor on image to zoom/Click text to open image
Figure 9: Flaps approximated with primary closure.



Cursor on image to zoom/Click text to open image
Figure 10: Panoramic X-ray after implant surgery.



Discussion

Diagnostic imaging techniques help develop and implement a comprehensive treatment plan for the implant team. Imaging modality can be organized into three phases. Phase one objectives include all necessary information to determine the quantity, quality, and angulation of bone; the relationships of critical structures to the prospective implant site; and the presence or absence of disease around the surgical site. Phase two objective is to evaluate the site of surgery during and immediately after surgery, assist in the optimal position and orientation of dental implant. Phase three is termed post-prosthetic implant imaging. Panoramic radiographs are sufficiently reliable to evaluate the available bone height before inserting posterior mandibular implants and are a widely used standard radiographic examination tool when planning an implant treatment because they impart a low radiation dose while giving the best radiographic survey [2] [3].

For a successful implant supported definitive restoration the implant must be placed at a correct pre-planned position and angulation [4]. The implant should be placed at least 1.5 mm from the adjacent teeth with a minimum 3 mm inter-implant distance. The distance of implant from buccal and lingual cortical plates should be greater than 0.5 mm. In the buccolingual plane, the angle between the implant trajectory and residual bone trajectory should be less than 20° to prevent unfavorable bending moment [5] [6] [7]. It has been well documented in literature that the implants placed using templates are more accurately positioned than those without the stent [8][9].

Mietner et al. (2003) stated that the simple implant template provides high reliability and accuracy. Yet simple device that can be used by anyone skilled in the art of dental surgery.

Template designs in the late 90's were either diagnostic or surgical and focused on determining the correlation of implant site with surrounding vital structures [10][11]. They did not help in determining the relative parallelism between adjacent implants and/or teeth [12].

This template does not require advanced computer technology or complicated equipment and complicated steps for fabrication. Significant time and cost savings can therefore be realized by using this template to perform dental implant osteotomy. It provides adequate accuracy as well in terms of implant position and angulation. It stands out from other techniques that it helps to determine relative parallelism between adjacent implants and/or teeth.


Conclusion

This case report has outlined the fabrication and clinical use of the diagnostic-surgical implant stent. Implant rehabilitation with use of stent provides greater precision. A simple planning can greatly improve the efficiency of implant therapy.


References
  1. Tsuchida F, Hosoi T, Imanaka M, Kobayashi K. A technique for making a diagnostic and surgical template. J Prosthet Dent 2004 Apr;91(4):395–7.   [CrossRef]   [Pubmed]    Back to citation no. 1
  2. Dula K, Mini R, van der Stelt PF, Buser D. The radiographic assessment of implant patients: decision-making criteria. Int J Oral Maxillofac Implants 2001 Jan-Feb;16(1):80–9.   [Pubmed]    Back to citation no. 2
  3. Harris D, Horner K, Gröndahl K, et al. E.A.O. guidelines for the use of diagnostic imaging in implant dentistry 2011. A consensus workshop organized by the European Association for Osseointegration at the Medical University of Warsaw. Clin Oral Implants Res 2012 Nov;23(11):1243–53.   [CrossRef]   [Pubmed]    Back to citation no. 3
  4. Lazzara RJ. Effect of implant position on implant restoration design. J Esthet Dent 1993;5(6):265–9.   [CrossRef]   [Pubmed]    Back to citation no. 4
  5. Almog DM, Torrado E, Meitner SW. Fabrication of imaging and surgical guides for dental implants. J Prosthet Dent 2001 May;85(5):504–8.   [CrossRef]   [Pubmed]    Back to citation no. 5
  6. Talwar N, Singh BP, Chand P, Pal US. Use of diagnostic and surgical stent: a simplified approach for implant placement. J Indian Prosthodont Soc 2010 Dec;10(4):234–9.   [CrossRef]   [Pubmed]    Back to citation no. 6
  7. Almog DM, Onufrak JM, Hebel K, Meitner SW. Comparison between planned prosthetic trajectory and residual bone trajectory using surgical guides and tomography--a pilot study. J Oral Implantol 1995;21(4):275–80.   [Pubmed]    Back to citation no. 7
  8. Engelman MJ, Sorensen JA, Moy P. Optimum placement of osseointegrated implants. J Prosthet Dent 1988 Apr;59(4):467–73.   [CrossRef]   [Pubmed]    Back to citation no. 8
  9. Modica F, Fava C, Benech A, Preti G. Radiologic-prosthetic planning of the surgical phase of the treatment of edentulism by osseointegrated implants: an in vitro study. J Prosthet Dent 1991 Apr;65(4):541–6.   [CrossRef]   [Pubmed]    Back to citation no. 9
  10. Pesun IJ, Gardner FM. Fabrication of a guide for radiographic evaluation and surgical placement of implants. J Prosthet Dent 1995 Jun;73(6):548–52.   [CrossRef]   [Pubmed]    Back to citation no. 10
  11. Basten CH. The use of radiopaque templates for predictable implant placement. Quintessence Int 1995 Sep;26(9):609–12.   [Pubmed]    Back to citation no. 11
  12. Takeshita F, Tokoshima T, Suetsugu T. A stent for presurgical evaluation of implant placement. J Prosthet Dent 1997 Jan;77(1):36–8.   [CrossRef]   [Pubmed]    Back to citation no. 12

[HTML Abstract]   [PDF Full Text]

Author Contributions
Abdul Habeeb Bin Mohsin – Conception and design, Acquisition of data, Analysis of data, Drafting the article, Critical revision of the article, Final approval of the version to be published
Husna Khan – Conception and design, Drafting the article, Final approval of the version to be published
Mohd. Aijaz Ahmed – Conception and design, Acquisition of data, Analysis of data, Drafting the article, Final approval of the version to be published
M. Mahesh Kumar – Conception and design, Acquisition of data, Critical revision of the article, Final approval of the version to be published
Sarah Samee – Acquisition of data, Analysis and interpretation of data, Drafting the article, Critical revision of the article, Final approval of the version to be published
Guarantor of submission
The corresponding author is the guarantor of submission.
Source of support
None
Conflict of interest
Authors declare no conflict of interest.
Copyright
© 2016 Abdul Habeeb Bin Mohsin et al. This article is distributed under the terms of Creative Commons Attribution License which permits unrestricted use, distribution and reproduction in any medium provided the original author(s) and original publisher are properly credited. Please see the copyright policy on the journal website for more information.